Lateral Force Calibration Method Used for Calibration of Atomic Force Microscope

Authors

  • Magdalena Ekwińska
  • Zygmunt Rymuza

DOI:

https://doi.org/10.26636/jtit.2009.4.985

Keywords:

AFM, calibration structure, cantilever, MEMS

Abstract

Modern heterogeneous micro- and nanostructures usually integrate modules fabricated using various materials and technologies. Moreover, it has to be emphasized that the macro and micro nanoscale material parameters are not the same. For this reason it has become crucial to identify the nanomechanical properties of the materials commonly used in micro- and nanostructure technology. One of such tests is a nanowear test performed using the atomic force microscope (AFM). However, to obtain quantitative measurement results a precision calibration step is necessary. In this paper a novel approach to calibration of lateral force acting on the tip of an AFM cantilever is discussed. Presented method is based on application of known lateral force directly on the tip using a special test structure. Such an approach allows for measurements of nanowear parameters (force, displacement) with the uncertainty better than ±3%. The calibration structure designed specifically for this calibration method is also presented.

Downloads

Download data is not yet available.

Downloads

Published

2009-12-30

Issue

Section

ARTICLES FROM THIS ISSUE

How to Cite

[1]
M. Ekwińska and Z. Rymuza, “Lateral Force Calibration Method Used for Calibration of Atomic Force Microscope”, JTIT, vol. 38, no. 4, pp. 83–87, Dec. 2009, doi: 10.26636/jtit.2009.4.985.